Chapter 4 SelfCalibration
Setting calibration schedule
Set the calibration schedule for SelfCalibration.
Attention
• Clock adjustment is necessary for SelfCalibration to operate.
Procedure
1.
Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Settings” from “SelfCalibration”, and press
.
3. Choose “Schedule” from “Settings”, and press
.
The schedule setting menu appears.
4. Use
or
to select the items to change, and press
.
Settings
Setting
Starting Time
Off
SelfCalibration is not executed.
Set the timing at which
SelfCalibration is executed.
Power Save
To be executed under any of the
following conditions.
• The monitor is in the power
saving mode or the power is
turned off when the time period
set in the schedule is reached
(“Type-1” or “Type- 2”).
• The monitor shifts to the power
saving mode or the power is
turned off when the time period
set in the schedule has elapsed.
Immediately
When the time period set in the
schedule has elapsed, it is executed
immediately.
Setting Type
*1
Type1
in months
Set according to the frequency at
which SelfCalibration is executed.
Type2
in weeks
Type3
in hours (monitor usage time)
Interval
Type1
1 - 12 months
Set the interval between the
completed SelfCalibration and the
next execution. The setting details
differ depending on the type of
“Setting Type” speci±ed.
Type2
1 - 5 weeks
Type3
50 - 500 hours
Week(s)
1st week - 5th week
Set which week to execute
SelfCalibration on for the month in
which calibration is executed.
Day
SUN/MON/TUE/WED/THU/FRI/SAT
Set which day to execute
SelfCalibration on for the week in
which calibration is executed.
Time
00:00 - 23:30
Set the time at which SelfCalibration
is executed.
*1
Settings that are used frequently can be preset.
5. Use
or
for adjustment.
6. Press
to exit.
The calibration schedule is set.
Note
• When the starting time is “±mmediately”, and the setting type is either “Type1” or “Type2”, SelfCalibration is
executed according to schedule even if the power is switched off using
.
• When the starting time is “Off”, the SelfCalibration menu shows “Setting incomplete.” However, SelfCalibration can
be executed manually.
Sample